No.102 GEER AGING OVEN

JIS A5756、B7757、C3005、K6257、K6723、K7212、K7368、ASTM D573、D5374、D5423、E145、UL-746B、IEC-60216-4-1、ISO 188、4577

No.102 GEER TYPE AGING OVEN

  • This tester is the most widely used heat aging tester for rubber and plastic.
  • Unlike ordinary heating ovens, the GEER AGING OVEN satisfies specifications indispensable to aging tests such as closed mechanism, uniform air velocity in the oven, ventilation rate measurement, and rotating specimen rack, which offers excellent reproducibility and reliable test results.

No.102 Specification

Specifications are subject to change without notice.

Model

Standard Type

Outsize Type

Oven Interior Dimensions

W450 × D450 × H500 mm W600 × D600 × H600 mm

Oven Volume

225,225 cm3 417,600 cm3

Temperature Range

Room temperature + 20–300℃

Average Air Speed

Low Speed Mode 0.5 ± 0.1 m/s , High Speed Mode 1.0 ± 0.2 m/s
(selected on touchscreen)

Air Exchange Rate

Calculated by the electric power consumption method

Air Exchange Rate Control

Adjustable within the following range
·3 to 10 times/hr (6.5 ± 3.5 times/hr)
·5 to 20 times/hr (12.5 ± 7.5 times/hr)
·60 times/hr or more (80 ± 20 times/hr)
·100 to 200 times/hr (150 ± 50 times/hr) *Optional

Sample Mounting Apparatus

Rotational Speed 7.5 ± 2.5 rpm

Safety Device

Overheat detection device, Earth leakage breaker

Accessories

Sample mounting apparatus (round, 2 holder frames),sample holding clips (60 pcs)

Option

UL Standard Specifications : UL standard (746B:2014) air exchange rate adjustable between 100-200times/hr, additional sample holding clips(30 pcs), CSV file output, square mesh shelf

Power

AC 200 V 3-phase 30 A 50/60 Hz,
For UL Standard Specifications:
AC 200 V 3-phase 40 A 50/60 Hz
*For No.103-U/L, a power source is required for each individual unit.

AC 200V, 3-phase, 40A, 50/60Hz
*For No.103-U/L, a power source is required for each individual unit.

Dimensions and Weight (Approx.)

SHF-S
W1,180×D700×H1,560 mm, 290 kg
SHF-SA
W1,250×D700×H1,650 mm, 300 kg

SHF-S
W1,330×D900×H1,610 mm, 365 kg
SHF-SA
W1,400×D900×H1,750 mm, 380 kg

Models

Specifications are subject to change without notice.
SHF-S

Includes a display of the air exchange rate measurement. The air dampers are opened and closed manually.

SHF-SA

The air dampers are opened and closed automatically to control the air exchange rate.

Inquiry Form

No.103-S-U/-S-L GEER AGING OVEN
(STACKABLE TYPE)

JIS A5756、B7757、C3005、K6257、K6723、K7212、K7368、ASTM D573、D5374、D5423、E145、UL-746B、IEC-60216-4-1、ISO 188、4577

No.103

This is a down-sized model of No.102 GEER AGING OVEN, making it stackable with the same functionality.
Necessary settings, such as average air velocity, can be configured via the touch panel. This model is the best solution for those wanting to utilize space.

*The chamber size is only available in the standard size (W450 × D450 × H500 mm).

Models

Upper Model: No.103-S-U / No.103-SA-U
Lower Model: No.103-S-L / No.103-SA-L
For details about the S-type/SA-type models, please refer to No.102.

Chamber Door Single Door

Oven Interior Dimensions

W450 × D450 × H500 mm
Oven Volume 207,658 cm3

Temperature Range

Room temperature + 20–300℃

Average Air Speed

Low Speed Mode 0.5 ± 0.1 m/s , High Speed Mode 1.0 ± 0.2 m/s
(selected on touchscreen)

Air Exchange Rate

Calculated by the electric power consumption method

Air Exchange Rate Control

Adjustable within the following range
·3 to 10 times/hr (6.5 ± 3.5 times/hr)
·5 to 20 times/hr (12.5 ± 7.5 times/hr)
·60 times/hr or more (80 ± 20 times/hr)
·100 to 200 times/hr (150 ± 50 times/hr) *Optional

Sample Mounting Apparatus

Rotational Speed 7.5 ± 2.5 rpm

Safety Devices

Overheat detection device, Earth leakage breaker

Accessories

Sample mounting apparatus (round, 2 holder frames),sample holding clips (60 pcs)

Options

UL Standard Specifications : UL standard (746B:2014) air exchange rate adjustable between 100-200 times/hr, additional sample holding clips(30 pcs), CSV file output, square mesh shelf

Power Source

AC 200 V 3-phase 30 A 50/60 Hz , For UL Standard Specifications: AC 200 V 3-phase 40 A 50/60 Hz
*For No.103-U/L, a power source is required for each individual unit.

Dimensions and Weight (Approx.)

W1,350(1,150)×D950×H950 mm, 290kg

* Specifications are subject to change without notice.
* The design of the product is subject to change without notice.

Inquiry Form

Contact Us

For over 70 years, Yasuda Seiki’s testing machines have supported quality control and research & development across various industries, meeting the diverse needs of our clients.