No.185 PARALLEL PLATE PLASTOMETER

No.185
PARALLEL PLATE PLASTOMETER

Overview of PARALLEL PLATE PLASTOMETER

This tester is to conduct heat deformation tests to rubber and plastic electric insulated wires.
Placing the test specimen between the parallel plates and adding prescript load and heat for 30 minutes, the operator is to measure the reduction rate in thickness of the test specimen before and after the test has been conducted.

Reference Standards

JIS C3005, C3307, K6723, ASTM D621

Specification

Specifications are subject to change without notice.
ModelNo.185

Number of Test Stations

1, 3, or 5 stations

Test Load

Max. 50 N (includes counterweight)

Displacement Measurement

Analog dial gauge graduation 1/100 mm, stroke length max. 10 mm
(option: digital dial gauge,stroke length max. 25 mm)

Temperature

Max. 200 °C

Anvils

Diameter ø40 mm (option: UL standard diameter ø3/4″ or ø3/8″)

Accessory

Metal rod (radius 5 mm)

Options

Test chamber with interior light

Power Source

AC 200 V 1 or 3 phase15–30 A 50/60 Hz

Dimensions and Weight (Approx.)

1 station: W780×D825×H1,800mm, 142kg
3 stations: W1,250×D1,120×H1,800mm, 286kg
(tabletop model: H952mm)


No.185-AUTO PARALLEL PLATE PLASTOMETER (AUTOMATIC)

No.185-AUTO
PARALLEL PLATE PLASTOMETER (AUTOMATIC)

Overview of PARALLEL PLATE PLASTOMETER (AUTOMATIC)

This tester is to conduct heat deformation tests to rubber and plastic electric insulated wires.
Placing the test specimen between the parallel plates and adding prescript load and heat for 30 minutes, the operator is to measure the reduction rate in thickness of the test specimen before and after the test has been conducted.
The operator is to set the pre-heating time and the testing time on the touch panel, and then the tester will load, unload, and show and print the test data automatically.

Reference Standards

JIS C3005, C3307, K6723, ASTM D621

Specification

Specifications are subject to change without notice.
ModelNo.185-AUTO

Number of Test Stations

3, or 5 stations

Test Load

Max. 50 N (includes counterweight)

Displacement Measurement

Digital contact sensor resolution 1/100 mm, stroke length max. 50 mm, touchscreen display

Temperature

Max. 200 °C

Anvils

Diameter ø40 mm (option: UL standard diameter ø3/4″ or ø3/8″)

Accessory

Metal rod (radius 5 mm)

Options

Test chamber with interior light, PC software

Power Source

AC 200 V 1 or 3 phase15–30 A 50/60 Hz

Dimensions and Weight (Approx.)

3 stations: W1,280×D880×H1,800mm, 343kg
5 stations: W1,420×D880×H1,800mm, 375kg